Subject: force degradation, lux hours, photo chamber
force degradation, lux hours, photo chamber

Check for Interference from force degradation study
Procedure
Blank, unstressed sample preparations were prepared as per the method and unstressed
placebo was prepared as per protocol.

The Placebo were subjected to following stress
conditions.

Acid degradation : The sample was refluxed in 1N Methanolic HCl for 1hour.

Base degradation : The sample was refluxed in 1N Methanolic NaOH for 1 hour.

Oxidation : The sample was heated in 3% H2O2 at 40°C for 1 hour

Thermal degradation: The sample was kept at 100°C up to 72 hours.

Photo degradation : The sample was subjected to 1.2 million lux hours in the photo
chamber.

thanks

 alk moderator

link 9.04.2009 10:34 
http://www.stabilitytestchamber.com/photo_stability_test_chambers.htm

то есть устойчивость к воздействию света мощностью/силой/количеством освещения/облучения 1.2 млн люкс/час

а здесь http://www.pribor.ru/print.php?id=1204 просто:
Время достижения ICH рекомендаций 1200000 люкс-час

 